Customization: | Available |
---|---|
After-sales Service: | Lifetime Online Technial Support |
Warranty: | 1year |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model
|
CIC-D120+
|
|
Autosampler
|
Standard configuration
|
Number of samples: 120 × 2mL (standard
configuration). Maximum injection volume: ≥ 500 μL. Injection methods (three types): full loop/partial loop/Micro injection Injection repeatability: full loop injection: RSD ≤ 0.4%. Partial loop injection: RSD ≤ 0.5%. Micro injection: RSD ≤ 1.0%. |
Flow Path
|
PEEK material
|
Chemically inert, metal-free PEEK flow
paths, compatible with aqueous eluents of pH 0-14 and reversed-phase solvents |
Pump
|
Type
|
High-pressure and low-pulse two-piston
tandem advection pump |
Maximum
Pressure |
42MPa
|
|
Flow Rate Range
|
0.001~9.999mL/min
|
|
Tipping load
|
60kN
|
|
Pressure Ripple
|
6750ml
|
|
Max. torque
|
Pressure Ripple
|
|
Pump
|
Flow Precision
|
≤0.1%
|
Flow Accuracy
|
Flow Accuracy
|
|
|
External Diameter of the
Tube |
1/16"
|
Valve
|
Contact Material of the
Rotor |
PEEK
|
Control Mode
|
By Stepper motor
|
|
Power Supply
|
24V (DC)
|
|
Column
Heater |
Operating Temperature
Range |
Ambient + 5 to 60 ºC
|
Controlling Temperature
Accuracy |
±0.1ºC
|
|
Temperature Stability
|
≤0.1ºC/h
|
|
Columns Supported
|
2, 3, 4 and 5 mm I.D.
Maximum length 250 mm analytical column with 50 mm guard column. |
|
Conductivity
Detector |
Type
|
Auto-range conductivity detector
|
Cell Volume
|
Cell Volume
|
|
Detection Range
|
0~50000μS/cm
|
|
Resolution
|
≤0.0020nS/cm
|
|
Output Voltage
|
Output Voltage
|
|
Electronic Noise
|
0.02nS
|
|
Baseline Noise
|
≤0.001μS/cm
|
|
Baseline Drift
|
≤0.01μ
|
|
Temperature Range
|
Ambient +5ºC~60ºC
|
|
Controlling Temperature
Accuracy |
±0.01ºC
|
|
Temperature
Compensation |
1.7 %/ºC
|
|
Maximum Pressure
|
10.0MPa
|
|
Instrument Linearity
|
≥0.999
|
|
Quantitative Repeatability
|
≤0.5%
|
|
Power Supply
|
150W
|
|
External Size(L*W*H*)
|
350*470*510(mm)
|
|
Net Weight(KG)
|
26
|
|
Gross Weight(KG)
|
32
|