• up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
  • up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
  • up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
  • up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
  • up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
  • up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument

up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument

After-sales Service: Lifetime Online Technial Support
Warranty: 1 Year
Type: Spectroscopic Oil Measuring Instrume
Display: Digital
Control: Computer Control
Weight: 0-100Kg
Customization:
Gold Member Since 2024

Suppliers with verified business licenses

Manufacturer/Factory
  • Overview
  • Product Overview
  • PRODUCT SPECIFICATIONS
  • SIMILAR PRODUCTS
  • ABOUT US
  • FAQ
Overview

Basic Info.

Model NO.
UP-1001
Power Source
AC220V
Customized Support
OEM
Measurement Range
0.02~800mg/L
Detection Limit
<0.02mg/L
Classification
Other
Wavenumber Scanning Range
3400 Cm-1~2400 Cm-1 (2941nm-4167nm)
Baseline Drift
1%/4h
Communication Interface
RS232, USB
Transport Package
Wood Box
Trademark
Haicoer
Origin
Shanghai, China
Production Capacity
5000 Set/Sets Per Month

Product Description

Product Overview

UP-1001 Infrared spectroscopic Oil measuring instrument

The infrared spectrophotometer utilizes oil substances at wave numbers 2930 cm-1 (stretching vibration of the C-H bond in the CH group), 2960 cm-1 (stretching vibration of the C-H bond in the CH3 group), and 3030 cm-1 (absorption at the stretching vibration band of the C-H bond in aromatic hydrocarbons) for qualitative and quantitative analysis using the absorption and transformation of spectral energy.
 
up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument

PRODUCT SPECIFICATIONS

Model
UP-1001
Wavenumber scanning range
3400 cm-1~2400 cm-1 (2941nm-4167nm)
Wavenumber repeatability
± 1 cm-1
Wavenumber accuracy
± 1 cm-1
Detection limit
<0.02mg/L
Measurement repeatability
<2%
Measurement accuracy
± 2%
Linear range of absorbance
0.0000~1.9999AU
Measurement range
0.02~800mg/L
Minimum detection concentration
0.002mg/L (water sample concentration)
Baseline drift
1%/4h
Error of different proportions
≤ 5%
Communication interface
RS232, USB
Display
0.1-inch tablet computer
Overall dimensions (mm)
540 * 246 * 160 (mm)
Power supply
220V ± 40V 50Hz ± 1H 40W
Weight (KG)
13
Temperature
5-35
Humidity
<80%

Product Features

1.Good stability: Adopting an integrated optical system, with short optical path, high energy, stable signal, and high signal-to-noise ratio. 2.Small drift: The detector collects both the signal when the light source emits light and the signal when the light source is turned off, achieving real-time automatic zero clearing. 3.Accurate positioning: Using the cosine principle for precise wavenumber positioning scanning, the wavenumber positioning accuracy is less than one wavenumber. 4.Small measurement error for different ratios: simulate the oil composition in water, and the error for measuring any component of standard oil is less than 5%, making the instrument truly serve the actual water sample. 5.Full spectrum measurement: Full wave number measurement and real-time display of the spectrum, allowing for both qualitative analysis and quantitative measurement. 6.Oil measurement specialized software: Oil measurement specialized software (already applied for software) integrates spectrum scanning, analysis, calculation, and storage, making operation easier. Equipped with self inspection and result judgment functions: prompt if the energy is abnormal, and also indicate the possible cause for troubleshooting reference. Equipped with software to determine whether the sample exceeds the standard prompt function. 7.Remote control: The instrument adopts a 10.1-inch Windoes10 tablet computer, which is embedded in the host instrument. The tablet can be flexibly removed to achieve remote control, and the host is reserved with an external computer communication control interface. 8.Communication method: Bluetooth, RS232 communication.

operational principle

The UP-1001 infrared oil analyzer uses the principle of similar phase solubility to extract oil from water into tetrachloroethylene. After oil water separation, the oil is measured, and the oil is measured after passing through a magnesium silicate adsorption column. Oil substances have absorption at wave numbers of 2930 cm-1 (stretching vibration of C-H bonds in CH groups), 2960 cm-1 (stretching vibration of C-H bonds in CH3 groups), and 3030 cm-1 (stretching vibration bands of C-H bonds in aromatic hydrocarbons). Qualitative and quantitative analysis is carried out using absorption and conversion of spectral energy.

application area

Suitable for the determination of industrial wastewater, domestic sewage, oil mist, petroleum in soil, and animal and plant oils.

Reference standards

HJ637-2018 Water Quality - Determination of Petroleum and Animal and Plant Oils - Infrared Spectrometric Method HJ1077-2019 Determination of Oil Smoke and Oil Mist in Exhaust Gas from Fixed Pollution Sources - Infrared Spectrometric Method HJ1051-2019 Determination of Petroleum in Soil - Infrared Spectrometric Method

application area

Suitable for the determination of industrial wastewater, domestic sewage, oil mist, petroleum in soil, and animal and plant oils.

SIMILAR PRODUCTS

up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument
 

ABOUT US

 
up-1001 Infrared Wastewater Measuring Instrument Oil and Grease Analyzer Infrared Spectroscopic Oil Measuring Instrument

FAQ

Q1What is your minimum order quantity?
A1The minimum order quantity is very low, and the minimum order is one set. For customized projects, it depends on different requirements.

Q2Do you offer guarantee for the products?
A2Yes, we offer 1-2 years warranty to our products.

Q3What is the delivrey time after an drder placed?
A3Our standard delivery is 3-5days when we have stock.We will let you know the lead time according to your oder quantity in time.

Q4What is the warranty?
A4The free warranty is one year from the day of Commissioning qualified.

Q5Can you tell me your main customers?
A5Sorry we could not tell you except customer's permission. At the same time,your information also will be protected.

Q6How can i get the after-service?
A6We will be responsible for our products in the valid time.
 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now